Flow assembly CYA680 | Endress Hauser E+H

Assembly for the Life Sciences and Chemical industries

Specs at a glance

  • Process temperature max.130°C 266°F
  • Process pressure Stainless steel: max. 16 bar; PVDF: max 4 bar

Field of application

The CYA680 flow assembly is the simple solution for pH and conductivity measurement in pipelines of hygienic and sterile applications. It is designed for inline cleaning and sterilization and perfectly suited for small pipe diameters, offering maximum safety for your high-quality products. For perfect performance in aggressive and corrosive media, CYA680 is available as PVDF version.

Benefits

  • Optimized product yield thanks to fast measured values in low volume
  • High product safety: Assembly is suitable for cleaning-in-place (CIP) and sterilization-in-place (SIP)
  • Compliant with strict Life Sciences regulations: Certified biocompatibility according to USP class VI, FDA-listed seals and hygienic electropolished surface Ra=0.38 µm (15 µinch)
  • Suitable even for aggressive and corrosive media with PVDF version

Spare parts & accessories

DescriptionCategoryOrder Code
1KIT CYA680 O-Rings set (KALREZ )further parts71321494
2KIT CYA680 O-Rings set (VITON )further parts71321495
3KIT CYA680 O-Rings set (EPDM )further parts71321496
4KIT CYA680 pH sensor adapter,120mm,316Lfurther parts71321499
5KIT CYA680 pH sensor adapter,120mm, PVDFfurther parts71323973
6KIT CYA680 cd sensor adapter, 120mm,316Lfurther parts71361303
7KIT CYA680 Dummy sensor 120 mm, 316L71321497
8KIT CYA680 Dummy sensor 120 mm, PVDF71321498